Plasmetrix Technologies has commercialized a side-illumination dark field microscopy platform that detects individual plasmonic nanoparticles on clinical specimens — enabling truly quantitative, multiplexed protein diagnostics from a simple adapter fitted to any standard pathology microscope. Developed at Polytechnique Montréal's LP2L laboratory by Prof. Michel Meunier and Dr. Sergiy Patskovsky, backed by a decade of peer-reviewed validation.
How lateral RGB illumination replaces the traditional condenser
Conventional dark field microscopy relies on a specialized condenser — typically a cardioid or paraboloid type — that projects a hollow cone of light through the specimen. Only scattered light enters the objective; direct illumination bypasses it. This imposes a hard constraint: the condenser NA must exceed the objective NA, forcing pathologists to use objectives with NA below ~1.0 or to stop down high-NA lenses. The result is degraded spatial resolution precisely when it matters most.
Plasmetrix's side-illumination (SI) approach takes an entirely different path. Light from arrays of discrete-wavelength RGB LEDs enters the specimen laterally — from the edge of the glass slide — rather than through a substage condenser. The light propagates within the glass by total internal reflection (TIR), governed by Snell's law at the glass–air interface, until it encounters refractive index discontinuities: tissue structures, cell membranes, or — critically — plasmonic nanoparticles. At these discontinuities, light scatters upward into the collection objective. Because no illumination beam ever enters the objective's acceptance cone, there is zero constraint on objective NA. Any standard air objective — including high-magnification 100× dry objectives — operates at full aperture.
The system operates in four steps. First, the SI adapter delivers spectrally controlled lateral illumination from the RGB LED array. Second, metallic nanoparticles undergo LSPR excitation and scatter at their characteristic wavelengths — blue for silver-rich alloys, green for Au/Ag compositions, red for gold-dominant particles. Third, the air objective collects only this scattered light against a dark background. Fourth, a CMOS camera captures the image and software decomposes the RGB channels to identify, count, and classify each nanoparticle.
Product configurations
Plasmetrix offers the SI adapter in five configurations: three for upright microscopes (automated software-controlled, manual, and Renishaw Raman-integrated) and two for inverted platforms (manual Nikon, motorized Nikon + Prior stage for whole-slide scanning). All share the same core optical design and mount at the specimen plane without modifying the condenser, filter turret, or objectives. View all configurations →
Technical advantages
Unrestricted objective NA
Any air objective at full aperture — no condenser matching. Oil-immersion excluded by design (preserves TIR), which pathologists prefer to avoid contamination.
Superior background suppression
TIR confines illumination within the glass slide, dramatically reducing autoscattering and dust artifacts versus transmission dark field.
Completely oil-free workflow
No oil anywhere in the optical path — prevents sample contamination, preserves specimens for re-imaging and archival storage.
Simplified alignment
No critical centering or focus adjustment. Produces uniform illumination across the entire field of view.
Inherent spectral selectivity
Discrete RGB LEDs enable color-resolved imaging without a spectrophotometer or tunable filter.
Full mode compatibility
No restriction on transmission, fluorescence, or phase contrast. Integrates with any inverted or upright microscope.
Conventional condenser DF
- NAcond > NAobj required
- Oil coupling to condenser
- Critical centering alignment
- Background leakage artifacts
- Sample contamination risk
Side-illumination DF (Plasmetrix)
- No NA constraint (air objectives)
- No oil anywhere — clean workflow
- Self-aligning adapter
- Intrinsic TIR background suppression
- No sample contamination
Publications
- 2025 Patskovsky, Perraton-Elorza, Largillière, Darviot, Harroun & Meunier — Methods and instrumentation for quantitative analysis with immunoplasmonic nanoprobes. Sensors & Actuators B, 139010
- 2024 Darviot et al. — Multiplexed immunolabelling of cancer using bioconjugated plasmonic gold–silver alloy nanoparticles. Nanoscale Advances
- 2023 Zapata-Farfan et al. — Dynamic multispectral detection of bacteria with nanoplasmonic markers. Nanoscale
- 2021 Marcoux-Valiquette et al. — Multiplexed plasmonic nano-labeling for bioimaging of cytological stained samples. Cancers 13, 3509 · PMC
- 2020 Nsamela Matombi et al. — Sensitive and rapid cancer diagnosis: toward fine-needle aspiration cytology. ACS Appl. Nano Mater.
- 2019 Qi, Darviot, Patskovsky & Meunier — Cost-effective side-illumination darkfield nanoplasmonic marker microscopy. Analyst 144, 1303
Why plasmonic nanoparticles outperform fluorescent labels
At LSPR resonance, the scattering cross-section scales with the sixth power of particle diameter (σscat ∝ d⁶). A 60 nm gold nanosphere scatters at approximately 800 nm² — roughly 800× larger than a comparable fluorescent bead. Unlike fluorophores, scattering labels do not photobleach, do not saturate, and signal scales linearly with illumination power.
By varying composition (Au, Ag, Au/Ag alloys) and shape (spheres, nanorods, bipyramids), a spectral palette spanning the visible range is engineered. Conjugated to antibodies, these become immunoplasmonic nanoprobes — individually resolved as bright, color-coded spots against the SI dark background.
Quantitative whole-slide integration
SI dark field microscopy with motorized stage scanning enables whole-slide imaging in dark field mode. Software detects, counts, and classifies every nanoparticle by RGB signature across thousands of fields, producing quantitative biomarker expression maps over entire tissue sections.
The binary readout — each nanoparticle either present or absent — eliminates the analog intensity ambiguities of chromogenic and fluorescence quantification. This technology could transform biomarker scoring from subjective visual assessment to objective, digital, multiplexed measurement — using nothing more than a simple adapter on microscopes pathologists already own.
Get the full white paper
Download the complete technical description with sample preparation protocols, optical detection methods, and the full referenced publication list.